|
Knowledge
of detecting spurious blood counts is key to effective
use of automated cell counters
Dhruv
Kamat
Though automated cell counters have become very popular,
the availability of validation criteria and reference
material is very less. Blind faith in automation should
be replaced by an understanding that automated blood
counts may be inaccurate, subject to conditions.
Also, the knowledge of detecting spurious blood counts
is the key to successful use of automation and so is
the subject of this article. It is the responsibility
of laboratory staff performing a counter authorising
report to detect inaccuracies whenever possible. All
suspected spurious cell counts/ morphologies should
be validated with microscopic examination of slides.
The validation of results generated by automated blood
cell counter requires:
(i) Knowledge that an instrument is capable of measuring
all variables accurately, has been correctly calibrated
and that quality control procedure indicates normal
functioning.
(ii) Assessment of each individual count as to whether
it is likely to be correct or that, alternatively, it
requires further review.
Some conditions that cause spurious counts and recommended
checks
Fault in specimen collection or storage
1. Blood from wrong patient blood specimen and request
relating to different patient.
2. Specimen diluted (e.g. taken from above intravenous
infusion or excess liquid EDTA relative to volume of
blood.)
3. Specimen taken in too high concentration of blood.
4. Presence of dust in locally prepared EDTA bottles
results in spuriously high cell counts on automated
cell counters.
5. Specimen haemoconcentrated due prolonged application
of tourniquet.
6. Specimen partly clotted/ lysed /aged blood specimen
inadvertently heated or frozen.
Faulty sampling
1. Failure in priming of instrument can cause spurious
high/ low counts.
2. Inadequate mixing or specimen of inadequate volume.
In general going by laws of fluid dynamics 60 inversions
of 2.5 ml blood specimen should be done prior to aspiration
on Automated cell counters. (A bulb mixer will be handy.)
3. Aspiration probe uncleaned or blocked by previous
sample. The cleanliness of probe can be visually checked
in between the runs.
4. Carry-over from preceding very abnormal specimen.(which
is minimal with modern instruments employing separate
dilution baths for primary and secondary dilutions.)
Faulty Calibration
1. Calibration performed without repeated successive
runs of calibrator. Minimum five runs of a calibrator
should be performed prior to deciding on calibration
factors.
2. Calibration performed under improper instrument conditions
or with improper calibrator.
All routine maintenance must be done thoroughly prior
to calibration.
Faulty maintenance /other instrument malfunctions/ reagent
failure
1. Contamination or deposition in counting fluidics.
Deposition in dilution baths and counting fluidics is
a common cause of random errors and should be kept an
eye upon.
2. Contaminated reagents. It is advised to keep the
reagent containers closed and away from dust.
3. Change in conductivity of reagents owing to high
temperature conditions.
4. Electrical noise in absence of proper earthling.
5. Mechanical or Hydraulic or Electronic failures inside
instrument.
Inaccuracy inherent in specific methodologies
Inaccuracies that are inherent to aperture size, electronic
editing, designing methodology of the instruments e.g.
Underestimation of MCV by impedance counters in presence
of Hypochromia or sometimes linearity problems associated
with higher Aperture size instruments.
Inaccuracy due to unusual specimen characteristics
Common cause is erronious HB or red cell indices caused
by cold agglutinins or due to unusual hyperlipidaemia
or factitious leukocytosis caused by insufficient lysing
of NRBCs in specimen characteristics neo-natal samples.
|